HOTHMAS-D5/10/30/50SZA Vickers Hardness Tester
Vickers Hardness Tester
- 测量分析系统HMAS‑硬度测量分析系统
- 光学成像系统维氏硬度大三通光学成像系统
HBS-3000MV · 2026-11-17 · +86 21 5748 2007
Brinell Hardness Tester
By submitting you agree to our privacy policy.
HBS-3000MV,是一款布氏硬度测试设备,采用全自动试验力加载模式,支持多种布氏硬度标尺测量,试验力级数丰富。它执行多项国家标准,确保测试规范性。设备配备
200W
工业相机,采用平板软件读数和自动识别压痕的数据采集方式,光源亮度可调,试验力保荷时间灵活设置,在结构参数上试件最大高度、压头中心到内壁距离等设计合理,适用于对金属材料布氏硬度的高精度检测,尤其在冶金、机械制造等领域的材料硬度质量控制与科研场景中,能满足对碳钢、合金钢、铜合金等材料的硬度与强度换算及精准检测需求。
产品特点:
型号
HBS‑3000MV
测量标尺
HBW2.5/62.5、HBW2.5/187.5、HBW5/62.5、HBW5/125、HBW5/250、HBW5/750、HBW10/100、HBW10/250、HBW10/500、HBW10/1000、HBW10/1500、HBW10/3000
试验力级数
62.5kgf
(612.5N)、100kgf(980N)、125kgf(1225N)、187.5kgf(1837.5N)、250kgf(2450N)、500kgf(4900N)、750kgf(7350N)、1000kgf(9800N)、1500kgf(14700N)、3000Kgf(29400N)
转换标尺
GB/T1172‑碳钢硬度与强度换算值;GB/T1172‑碳钢及合金钢硬度与强度换算值;GB/ASTM/ISGB
3771‑1983
铜合金硬度与强度换算值
试验力加载
自动(加载/保荷/卸载)
执行国家标准
GB/T231.1
国家标准,GB/T231.2
国家标准,JJG150
检定规程
工业相机
200W
理论硬度测量范围
8‑650HBW
硬度值显示方式
平板(11.6
寸)软件读数
光源亮度调节
0‑10 级可调
测量数据采集方式
自动识别压痕
试验力保荷时间
0‑99 秒,最小变动量
1 秒
试件最大高度
230mm
压头中心到内壁距离
150mm
试台直径
大平台 φ200mm、中平台
φ77mm、V
型台 φ77mm
电源电压
AC 220V/50Hz
外形尺寸
512*223*745mm(长
* 宽 * 高)
净重
120.5kg
毛重
146.0kg
HBS-3000MV Brinell Hardness Tester supplied by CHWES. 0 key specifications available on request.
* Generated by rule engine over the product database; for pre-selection only.
Hardness testing presses an indenter into the surface under a constant load and converts indentation size or depth into a hardness value. Vickers uses a 136° diamond pyramid: HV = 0.1891 × F(N)/d²(mm), where d is the mean diagonal.
Hot hardness is performed in vacuum or high-purity inert gas. Specimen and indenter are heated independently and must be at the same temperature or corrected per the standard; loading, dwell and unloading all happen inside the hot zone, and the indentation is measured through a high-temperature optic or after cooling. The HMAS series uses a uniform hot zone with ±5 °C uniformity from 200 to 1600 °C.
1. Rockwell: HR = K − h/0.002 (K = 100 for HRC, 130 for HRB). At h = 0.078 mm, HRC = 61.
2. Brinell: HBW = 2F/(πD(D−√(D²−d²))); with F=1838.6 N, D=2.5 mm, d=1.1 mm, HBW ≈ 192.
3. Vickers: HV = 0.1891·F/d² (F in N); F=9.807 N with d=0.05 mm gives HV ≈ 742.
4. Test forces: choose Brinell loads by 0.045F/D²; Rockwell 10 kgf preliminary with 60/100/150 kgf total; surface scales 3 / 15/30/45 kgf.
5. Conversions: HRC ≈ HBW/10 only as a rough mid-range guide — official reports must follow the GB/T 1172 tables.
| Standard | Title | Scope |
|---|---|---|
| GB/T 4340.1-2009 | Vickers hardness test — test method | Vickers |
| GB/T 230.1-2018 | Rockwell hardness test | Rockwell |
| GB/T 231.1-2018 | Brinell hardness test | Brinell |
| ISO 6507-1 | Vickers hardness test | Export |
| ASTM E92 | Vickers hardness of metallic materials | Americas |
| GB/T 17394.1-2014 | Leeb hardness test | Portable |
Hot hardness has no unified national standard; ASTM E92 practice is normally cited with temperature, atmosphere, dwell and indenter material stated in the report.
1. Principle follows the target: Rockwell for macro hardness and heat-treatment acceptance; Brinell for cast iron, annealed and forged parts where a large indent is representative; Vickers or micro-Vickers for thin layers, coatings and small areas.
2. Bench or portable: bench units (HRC-150SMJ, HBRVS-187.5) where parts can be brought in; portable Leeb for large moulds, in-service piping and non-removable parts, corrected against a bench reference, typically within ±2 HRC.
3. One for many: mixed, low-volume shops do well with a Brinell-Vickers-Rockwell combination unit covering HRA/HRB/HRC/HV/HB.
4. Automation: manual below 100 parts per operator per day; motorised elevation, automatic loading and data upload (HRC-150SMJ) above 200 or for unattended runs.
5. Reference blocks: two to three blocks covering the working range, checked every shift and sent for annual calibration.
6. Specimen requirements: Rockwell needs thickness at least ten times the indent depth and Ra ≤1.6 µm; otherwise switch to Vickers or superficial Rockwell.
| Model | Principle | Range | Typical use |
|---|---|---|---|
| HRC-150SMJ | Rockwell, motorised digital | HRA/HRB/HRC | Batch sampling in heat treatment |
| HRC-150SM / HRC-150M | Rockwell | HRA/HRB/HRC | Incoming and in-process checks |
| HRC-45M | Superficial Rockwell | 15N/30N/45N | Sheet, plating, case layers |
| HRS-150M / XHRS-150M | Digital Rockwell | HRA/HRB/HRC | Lines needing data upload |
| HBRVS-187.5 | Brinell-Vickers-Rockwell | HBW/HRA/HRB/HRC/HV | Mixed low-volume production |
Three rules for trustworthy data: check the reference block every shift and record it; keep the support face flat and clean; keep indent spacing at least three indent diameters.
Specimen. Test and support faces parallel; thickness at least 1.5 × diagonal (Vickers) or 10 × depth (Rockwell); Ra ≤ 0.4 µm, free of scale and oil.
Room temperature. Select load, apply within 10–15 s, dwell 10–15 s, unload, measure diagonals. Spacing between indentations ≥3d, from the edge ≥2.5d.
Hot hardness. Load the specimen, evacuate or backfill argon, ramp and soak 20–30 min, lower the indenter, load and dwell, unload, cool or measure hot, then apply the temperature correction.
Gradient. Step every 0.05–0.1 mm to the hardness plateau; determine effective case depth at 550 HV1 per GB/T 9450.
Quality control. Check a certified block (e.g. 700 HV1) each shift, recalibrate beyond ±3%, and log furnace curves and atmosphere for hot tests.