Food testing2024

TY-9620 GFAAS for Cadmium in Rice

TY-9620 graphite-furnace AAS measures cadmium in rice per GB 5009.15 with a 0.001 mg/kg detection limit, deviating <5% from ICP-MS; 120 samples daily during harvest with 100% interception of over-limit grain.

회사명
A grain and oil quality centre, C. China
주소
Changsha, Hunan
TY-9620 GFAAS for Cadmium in Rice
0.001 mg/kgCd detection limit
35 元Cost per sample

The centre screens incoming rice for cadmium against the 0.2 mg/kg national limit. Everything went to ICP-MS - three days and 200 yuan per sample - while 120 truckloads a day waited at the weighbridge during harvest.

Solution: a TY-9620 graphite-furnace AAS, measuring after microwave digestion per GB 5009.15 with ammonium dihydrogen phosphate as matrix modifier, reaching a 0.001 mg/kg Cd detection limit. Parallel samples in the 0.1-0.4 mg/kg range were benchmarked against ICP-MS.

Results: deviation under 5% with 92-104% spike recoveries. Throughput reached 120 samples a day with 2-hour turnaround, intercepting 100% of over-limit grain; per-sample cost fell to 35 yuan, saving about 780000 yuan in one harvest season.