The centre screens incoming rice for cadmium against the 0.2 mg/kg national limit. Everything went to ICP-MS - three days and 200 yuan per sample - while 120 truckloads a day waited at the weighbridge during harvest.
Solution: a TY-9620 graphite-furnace AAS, measuring after microwave digestion per GB 5009.15 with ammonium dihydrogen phosphate as matrix modifier, reaching a 0.001 mg/kg Cd detection limit. Parallel samples in the 0.1-0.4 mg/kg range were benchmarked against ICP-MS.
Results: deviation under 5% with 92-104% spike recoveries. Throughput reached 120 samples a day with 2-hour turnaround, intercepting 100% of over-limit grain; per-sample cost fell to 35 yuan, saving about 780000 yuan in one harvest season.