Photoelectric Autocollimators

0.1″ class, 10–50 m

Self-developed HXJL/HXM/HXDS photoelectric autocollimators with 0.1–1″ resolution and 10–50 m range, in single-axis, dual-axis and large-field versions, for straightness, flatness and precision angle metrology.

0.1″ class, 10–50 m

Photoelectric Autocollimators · 73 products

01

Working principle

Photoelectric Autocollimators

An autocollimator projects an illuminated cross-line as a parallel beam onto an external mirror. The mirror returns the beam, which re-images on the reticle or a CCD/CMOS sensor. If the mirror tilts by θ, the return image shifts by s = 2f·tanθ — the angle is doubled optically and magnified by the focal length.

Photoelectric readout (CCD/CMOS centroiding or PSD differential detection with software subdivision) reaches 0.1″ resolution. Dual-axis models read horizontal and vertical tilt simultaneously; large-field versions such as the HXDS15-4000 (4000×3000″) image a wide angular range for dynamic measurements.

Stepping a mirror bridge along a guideway converts per-step tilts θi into elevation differences hi = L·tanθi for straightness and flatness; with polygons or multi-tooth tables the same instrument calibrates circular indexing.

02

Process animation

Photoelectric Autocollimators

  1. Setup: mount the mirror on a bridge of span L at the start of the guideway.
  2. Zeroing: adjust the autocollimator until the cross image centres, record the initial reading.
  3. Step: advance the bridge by L each station, logging θi (both axes on dual-axis models).
  4. Evaluate: accumulate hi = L·tanθi and rate straightness or flatness by end-point or minimum-zone methods.
  5. Indexing: with a polygon, successive face readings give the index error; issue the report.
0102030405
03

Calculation & formulas

Photoelectric Autocollimators

1. Angle-image relation: θ = s/(2f); with f = 500 mm and s = 5 µm, θ ≈ 1.03″ (1″ ≈ 4.85 µrad).

2. Resolution: 0.1″ corresponds to s ≈ 0.485 µm at f = 500 mm — pixels ≤5 µm plus 1/10 subdivision.

3. Elevation step: h = L·tanθ; a 100 mm bridge at 2″ gives h ≈ 0.97 µm.

4. End-point straightness: cumulative deviations Δi = Σhk − (i/n)Σh, error f = max(Δ) − min(Δ); the minimum-zone method rotates the datum per convex/concave/alternating criteria.

5. Indexing: per-face errors φj minus nominal; the closure Σφ = 0 checks the full circle.

04

Standards

Photoelectric Autocollimators

StandardTitleScope
JJG 202-2007Verification regulation of autocollimatorsIndication error and resolution
GB/T 11336-2004Detection of straightness errorGuideway and surface-plate straightness
GB/T 1958-2017GPS — detection and evaluation of geometrical deviationsFlatness, squareness and other form errors
ISO 230-1Machine tools — geometric accuracy testsMachine acceptance
JJG 283-2007Multi-tooth indexing table verificationCircular indexing calibration

Accredited labs must also hold 20±0.5 °C and vibration isolation, with instrument error within the 0.5″/1″ class.

05

Selection guide

Photoelectric Autocollimators

1. Resolution decides acceptance capability: 0.1″ (HXJL300-0.1, HXJL300U-0.1, HXJL300UH-0.1) for precision machines, rotary tables and laboratory standards; 0.2–0.3″ for general machine acceptance; 0.5–1″ (HXM series) for on-site alignment and long-range monitoring.

2. Distance decides aperture: 5–10 m with HXM2010/HXJL200; 20–30 m with HXM2020/HXM3030; 35–40 m with HXM3540; 50 m baseline with HXM4050M-1. Mirror aperture and light budget matter more as distance grows.

3. Single or dual axis: single-axis HXJL for one-direction straightness; dual-axis HXM for both directions and polygon indexing.

4. Dynamic work: HXDS15-4000 with a large field and high sampling rate for rotary-table oscillation, shaker tables and thermal drift.

5. Accessories: mirror bridges at 50/100/200 mm span, 12/24/36-face polygons, pentaprism for squareness, intermediate targets for long runs, thermal hoods for workshops.

6. Software and compliance: end-point, minimum-zone and least-squares evaluation with ISO 230-1 reports; labs also need the standard-artefact interface for verification.

ModelResolutionDistanceTypical use
HXJL300UH-0.10.1″0–10 mPrecision tables, lab standards
HXJL300U-0.1 / HXJL300-0.10.1″0–10 mPrecision guideways and plates
HXJL300-0.2 / -0.30.2–0.3″0–10 mGeneral machine acceptance
HXM2010 / HXM20200.5–1″10–20 mMid-size gantry, moulding machines
HXM3030 / HXM35400.5–1″30–40 mLarge gantry, marine parts
HXM4050M-11″50 mLong guideways, cross-shop datums
HXDS15-4000Dynamic dual-axisLarge fieldTable oscillation, vibration

Common mistakes: using a 1″ unit to trend a 0.02 mm/m guideway; ignoring bridge sag under mirror weight; not controlling air currents and floor vibration, so readings never settle.

06

Test method

Photoelectric Autocollimators

Environment: 20±0.5 °C (±2 °C on the floor with compensation), shield air currents and sunlight; soak the instrument and bridge for 30 minutes on an isolated base.

Straightness procedure: ① clean the guideway and set the bridge span L; ② zero at the first station after 10 minutes of warm-up; ③ step by exactly L (within 0.5 mm), logging both axes; ④ evaluate by end-point and minimum-zone methods and report the larger; ⑤ record span, temperature and uncertainty (typically U ≈ 0.5″ + 2% of reading).

Intermediate checks: monthly 12-face polygon closure (Σ = 0, residual ≤0.3″); send standards out every two years.

Pitfalls: span errors feed straight into height differences; air currents jitter the centroid; thermal gradients bend long guideways optically — measure morning and afternoon and average.

07

Industry applications

Photoelectric Autocollimators

  • Machine tools: gantry and boring-mill guideway straightness, column and spindle squareness, table flatness to ISO 230-1 and GB/T 11336; repeatability ≤1 µm over 24 m with the end-point method.
  • Metrology institutes: surface-plate flatness, cross-checks with electronic levels, small-angle standard build-up; indication error within ±0.5″/±1″ per JJG 202-2007.
  • Indexing and rotary tables: multi-tooth tables, optical and CNC rotary tables, RV reducer backlash; full-circle checks with 12/24/36-face polygons, indexing error ≤±1″.
  • Semiconductor stages: micro-angle feedback and thermal drift monitoring on air-bearing and wafer stages, resolution ≤0.01″ at ≥10 Hz.
  • Inertial navigation and optics: reference-mirror alignment and angle transfer for gyro platforms, theodolites and interferometers.
  • Energy equipment: turbine bearing alignment, wind-turbine flange squareness, long hydro shaft lines.
  • Rail and elevators: guide-rail straightness and twist, body datum flatness, often at 30–50 m working distance.
  • Research and teaching: optical bench levelling and micro-angle measurement.
Industry cases

Case Studies

View all→

Price and lead time for this model

Our engineers reply within one business day with selection advice, acceptance criteria and lead time.