The lab serves connector and wire-harness makers for RoHS compliance. Outsourced ICP testing took five days at high per-sample cost, could not meet urgent demand, and was losing about 30% of orders a year.
Solution: an ICPE-9000 simultaneous ICP-OES with microwave digestion, measuring Cd, Pb, Hg and total Cr per IEC 62321-5. Axial viewing gives a 0.2 mg/kg Cd detection limit, with matrix-matched spike recoveries held at 95-105%.
Results: 60 samples a day with same-day reports; certified reference material at 10 mg/kg Cd reads within 3% deviation, passing CNAS proficiency testing. New testing revenue of about 1.5 million yuan a year, with urgent-order loss eliminated.