What can a photoelectric autocollimator measure?

Micro-angles down to 0.01″. With accessories it covers straightness, flatness, parallelism, squareness and machine-tool geometry. Unlike interferometers it measures angle rather than displacement.

Resolution or accuracy — which matters more?

Resolution is only the display step. What matters is the indication error over the full range and repeatability. Check full-range error, local accuracy within ±10″, and 10-run repeatability.

Does accuracy drop with longer working distance?

Yes. Return intensity falls with the square of distance and the spot grows, raising centroid noise; air turbulence adds jitter. Use longer focal length and larger mirrors for long range.

Why is a precision flat mirror necessary?

The instrument measures the mirror normal, so mirror figure error adds directly to the result. Use λ/10 or better flats with fine pitch/yaw adjustment.

How is guideway straightness measured with an autocollimator?

Mount a mirror on the moving carriage, read angle at equal intervals, integrate angle against travel and evaluate the curve by the minimum-zone method. Measure both directions to cancel backlash.