The institute previously extrapolated hot hardness from room-temperature data, causing over 25% deviation between blade life models and rig tests.
Solution: IMAS-HT1500CM with a 10-3 Pa vacuum chamber applied 1 kgf Vickers indents on 5 sections (root to airfoil) at 900/1000/1100℃, with 20 min soaking and in-situ 10X optical reading per GB/T 4340.1.
Result: airfoil hardness fell from 468 HV at room temperature to 285 HV at 1100℃ (-39%), with the γ′ dissolution knee near 1050℃. The life-model deviation converged to within 8% after adopting the measured curve.