The institute organized the first national 1200℃ interlaboratory comparison of Vickers hardness and needed a pilot instrument with proven long-term repeatability plus homogeneous transfer blocks.
Solution: HMAS-HT1200 verified the homogeneity of nickel-alloy transfer blocks at 1200℃ (10-indext HV10 range ≤2.5%); the comparison followed JJG 112 and ISO 6507-2 with 10 kgf load, 15 s dwell, and argon protection.
Result: all six participating labs stayed within ±4% of the reference value with En < 1. The report now underpins the national traceability chain for hot hardness.