Clip-on extensometers left knife-edge marks on CFRP and slipped, so strain readings jumped and modulus scatter across five specimens reached ±6%, making it impossible to separate material variation from measurement error.
Solution: WDS-100G electromechanical frame (100 kN, ±0.5% speed accuracy) with a video extensometer using target marks at 0.5 µm resolution; loaded at 2 mm/min per ASTM D3039 with modulus fitted between 0.1% and 0.3% strain.
Result: modulus repeatability ±1.8% with undamaged specimens available for fractography; a 4% difference between two batches traced to fibre volume fraction in the prepreg, tightening incoming inspection.