The wafer inspection stage must hold angle stability better than 0.05″ while moving, yet images lost focus slightly after four hours, suggesting thermal drift of the air-bearing stage. The existing electronic level resolved only 1″, far too coarse.
Solution: HXJL300UH-0.1 at 0.01″ resolution aimed at the stage datum mirror, logging at 10 Hz for eight hours together with ambient and air-supply temperatures, then correlating angle and temperature traces.
Result: pitch drift 0.11″ and yaw 0.06″ per four hours, strongly correlated with a 0.8 °C supply-air rise (R²=0.94); tightening air-supply control to ±0.1 °C reduced drift to 0.03″ and eliminated the defocus.