Indexing error of the A/C axes directly affects impeller and mould contour accuracy. The previous dual-frequency laser with a rotary attachment needed two hours to set up and six hours for 15 positions, and any floor vibration forced a restart.
Solution: HXM2020-0.5 dual-axis autocollimator with a 24-face polygon calibrated by a higher-level institute (U=0.3″). Every 15° of table rotation the polygon face returns the beam and the angle deviation is read directly; the software evaluates by minimum zone and plots the full-circle error curve.
Result: 70 minutes per table with repeatability within 0.8″; a 4.5″ periodic error near 180° on the C axis was traced to worm-wheel eccentricity and reduced to 1.2″ after rework. The procedure is now a mandatory final-inspection step.