Metrology and testing2022

0.1 arcsec traceability for angle gauges and polygons

YR-0.1S serves as the primary standard for angle traceability, calibrating grade-0 angle gauges and polygons per JJG 70 for 200+ labs yearly.

Company
A municipal metrology institute, S. China
Address
Guangdong
0.1 arcsec traceability for angle gauges and polygons
U=0.15″ (k=2)Measurement uncertainty

The institute’s 0.2 arcsec autocollimator could not meet the uncertainty budget for grade-0 angle gauges (±1 arcsec limit), restricting its angle traceability business.

Solution: YR-0.1S (0.01 arcsec resolution, ±0.1 arcsec MPE) combined with a precision indexing table forms the calibration rig; the four-position method per JJG 70 removes polygon eccentricity, with evaluated uncertainty U=0.15 arcsec (k=2).

Result: the facility passed the national standard assessment and now calibrates 400+ grade-0 gauges and polygons yearly for more than 200 corporate labs across the province.