Difference between Ra, Rz and Rq?

Ra is the arithmetic mean, Rz the maximum height, Rq the RMS. Ra hides isolated deep valleys — add Rz for sealing, coating or fatigue surfaces.

How to choose cut-off and sampling length?

Per ISO 4288: 0.08 mm for Ra 0.006–0.02 µm, 0.25 mm for 0.02–0.1, 0.8 mm for 0.1–2, 2.5 mm for 2–10, 8 mm above. A wrong cut-off can double the result.

Profile meter vs roughness tester?

Roughness testers evaluate micro-irregularities; profile meters measure macro geometry such as radii, angles and groove depth. Combined systems give both in one trace.

What determines roundness instrument accuracy?

Spindle radial error (typically 0.02–0.05 µm) and sensor resolution, plus centring/levelling, filtering and evaluation method. Centre the part within 1/10 of tolerance.

Does measurement direction matter?

Yes — surfaces are anisotropic. Measure across the lay for turned surfaces, keep away from edges and keyways by at least 5× cut-off, and report 3 locations or more.