Bài kỹ thuật898 lượt xem

Hiệu chuẩn trực giao cho ống tự chuẩn trực hai trục

Trục X và Y không trực giao gây nhiễu chéo; cần hiệu chuẩn xoay và bù ma trận.

华旋光电技术部
Hiệu chuẩn trực giao cho ống tự chuẩn trực hai trục

Where crosstalk comes from

A dual-axis photoelectric autocollimator measures spot displacement in X and Y simultaneously on one position-sensitive detector, a PSD or CCD. Ideally the two measuring axes are orthogonal and aligned with the pitch and yaw directions of the reflector. In practice three deviations exist: rotation of the detector pixel array itself, azimuth misalignment between the optical axis and the reference face after mounting, and coupling between the mirror mount and the guideway attitude. As a result a pure pitch change produces a small false yaw reading and vice versa, with typical crosstalk of 0.5 to 2 percent of the coupled axis.

Impact assessment

On a 1 m guideway showing 10 arc-seconds of pitch and 10 arc-seconds of yaw, 1 percent crosstalk injects a 0.1 arc-second false component into each axis — exactly the resolution of a 0.1 arc-second instrument and therefore not negligible. In simultaneous dual-axis flatness grid measurement, crosstalk couples and amplifies along rows and columns, producing a saddle-shaped artifact in the evaluated surface.

Rotation-method calibration

The calibration uses a single-axis standard input: mount the reflector on a precision rotary stage, apply a standard angle sequence of ±50 arc-seconds about one axis only, for example pitch, and record the readings of the other axis; the slope is the crosstalk coefficient k_yx. Repeat with axes exchanged to get k_xy. The system is then described by a 2×2 coupling matrix M=[[1,k_xy],[k_yx,1]], whose inverse decouples live readings in software. Calibrate at each working distance, such as 1, 3 and 5 m, because the coefficients drift slightly with path length.

Acceptance and interval

After compensation, residual crosstalk should be below half the instrument resolution, i.e. 0.05 arc-seconds or less for a 0.1 arc-second class unit, corresponding to an orthogonality error within 0.5 arc-seconds. Calibrate annually, and always after shock or detector disassembly. A quick field check: measure the same angle before and after rotating the reflector 90 degrees about the vertical axis; the two axes should swap readings consistently, and any excess deviation calls for recalibration.

⇄0So sánhSo sánh thông số