Verification of Testing Machines per JJG 139: Force Error, Repeatability and Alignment

Class 1 machines require ±1.0 % force error and 1.0 % repeatability, verified at five or more points with a 0.3-class dynamometer; alignment is judged by bending strain within 5 %.

华旋光电技术部
Verification of Testing Machines per JJG 139: Force Error, Repeatability and Alignment

Verification items and limits

JJG 139 specifies the verification of tension, compression and universal testing machines: relative indication error, repeatability, relative zero error, resolution and safety devices. A class 1 machine, the mainstream for materials testing, requires force indication error within ±1.0 percent, repeatability within 1.0 percent, zero error within ±0.05 percent, and force stability over 30 s within ±1.0 percent. Class 0.5 machines halve these limits. Verification conditions are 10-35 °C and up to 80 percent relative humidity.

Standards and point layout

Use a 0.3-class proving instrument for class 1 machines, 0.1 class for 0.5-class machines, covering the verified range. Take at least five points evenly spread over 20 to 100 percent of the working range, three increasing-force runs per point. Relative error q=(F̄-F)/F×100 percent, repeatability b=(Fmax-Fmin)/F×100 percent, with F the reference force. Points below 20 percent of range are not verified — so never size a machine such that routine forces fall into the bottom quarter of its range.

Alignment assessment

Poor alignment adds bending to the specimen: yield strength reads high, elongation low, fatigue life collapses. JJG 139 assesses alignment with a dedicated alignment bar or strain-gauged specimen, and engineering practice quotes ASTM E1012: maximum bending strain within 5 percent of axial strain. Worn hydraulic grips and loose crosshead guides are the usual culprits.

Interval and maintenance

The verification interval is normally one year; reverify after overhaul, relocation or sensor replacement. Between verifications run intermediate checks with a working standard, such as a reserved proving ring or a validated specimen set. After software upgrades, confirm the force acquisition chain is unaffected — one known case failed dynamic indication because a filtering parameter changed. The uncertainty on the certificate must feed the uncertainty budget of your test reports.