How to Choose an Extensometer: Gauge Length, Travel, Accuracy Class and Contact Style

Rp0.2 needs a class 0.5 extensometer with gauge-length error within ±1 %; large deformation wants long travel, brittle materials want short gauge and fine resolution.

华旋光电技术部
How to Choose an Extensometer: Gauge Length, Travel, Accuracy Class and Contact Style

Accuracy class requirements

GB/T 12160, corresponding to ISO 9513, grades extensometers as 0.2, 0.5, 1 and so on, where the class is the relative indication error: class 0.5 means ±0.5 percent or ±1.5 μm, whichever is greater. Measuring Rp0.2 needs accurate elastic slope and tiny plastic strain, so the standard calls for class 0.5 or better; class 2 suffices for total extension at maximum force Agt. The gauge length itself must be within ±1 percent — ±0.5 mm on a 50 mm gauge — checked with a gauge ruler before mounting.

Gauge length and travel

Gauge length should match the parallel section and stay shorter than it: 50 or 25 mm for round bars, 50/80 mm for sheets. Travel decides how far into the curve you can follow: elastic plus yield needs only ±5 percent; elongation after fracture or uniform elongation wants 25 percent or more, or a scheme that removes the extensometer before maximum force and hands over to crosshead displacement. Highly deformable materials such as rubber use long-travel units reaching 500 mm on a 25 mm gauge.

Contact styles

Knife-edge contact is the general-purpose style; edge pressure must be moderate — too tight dents thin sheet, too loose slips. Fine wires need low-force clips. High-temperature work uses ceramic-rod extensometers rated to 1200 °C. For brittle materials, tiny specimens or scratch-sensitive surfaces, video extensometers avoid contact and added mass, though they depend on marking quality and ambient light; typical accuracy equals class 1, with high-end models reaching 0.5.

Routine checks

Extensometers are measuring instruments verified per JJG 762 on a one-year cycle. For daily checks use a displacement calibrator or dial-gauge fixture at 20, 50 and 100 percent of range; out-of-tolerance readings take the unit out of service. Before testing, run a preload to 10 percent and unload: zero drift should stay below 0.05 percent, otherwise check knife-edge seating and grip slip.