Hỏi đáp

Hỏi đáp

Hướng dẫn chọn máy · Phương pháp kiểm tra

How to read CMM accuracy specs?

MPEE = (1.8 + L/300) µm grows with length; MPEP is the probing error. Both matter — a poor MPEP ruins small-feature measurement.

Why does a CMM need a temperature-controlled room?

Steel expands ~1.2 µm per 100 mm per °C — the same order as machine accuracy. Specs assume 20±2°C and gradients within 1°C/h.

Touch trigger, scanning or optical probes?

Trigger probes for points and holes; scanning for free-form surfaces; optical for soft or thin parts. Multisensor probes finish everything in one setup.

How to keep measurement consistent?

Fix the alignment convention, point count and distribution, and evaluation algorithm, and document them. Differences of tens of microns between programs are common.

What routine maintenance is required?

Daily: clean guideways and probe, check air pressure, home the axes. Weekly: check scale heads and air bearing gaps. Yearly: calibrate and replace filters.

⇄0So sánhSo sánh thông số