Коррекция дисторсии объектива измерительных видеосистем

Даже телецентрические объективы имеют дисторсию; после калибровки по точечной мишени остаток <0,1 пикселя.

华旋光电技术部
Коррекция дисторсии объектива измерительных видеосистем

Sources and model of distortion

Lens distortion shifts image points near the field edge away from ideal positions, converting directly into dimensional error. Radial distortion follows the polynomial r'=r(1+k₁r²+k₂r⁴+k₃r⁶), with positive k₁ giving pincushion and negative giving barrel form; assembly decentration adds tangential terms p₁ and p₂ of the Brown model. A bi-telecentric lens typically distorts less than 0.05 percent, yet at the edge of an 8 mm field that is still a 2 μm shift — not negligible for machines rated at ±(3+L/200) μm.

Calibration procedure

Use a glass dot-array target with dot spacing accurate to ±1 μm, for example a 10×10 array at 1 mm pitch, covering the full field. Shoot the target at five to nine positions, locate dot centroids to 1/50 pixel, register against the nominal target coordinates, and fit the distortion coefficients. Pixel scale in μm per pixel is solved together with the coefficients in one least-squares run. The software then undistorts every frame before measuring.

Acceptance criteria

Verify residuals by measuring dot spacing with the target placed at nine grid positions of the field and comparing with nominal values. Requirements: residual distortion of 0.1 pixel or less (0.5 μm on a 5 μm/pixel system), and any two-point distance within the machine MPE. Recalibrate after changing lens, magnification or camera mounting. The target itself needs a traceable certificate with a one to two year cycle.

Common field problems

First, calibrating only at the field center while ignoring the edges, where distortion is largest. Second, target contamination drifting the centroids — clean with anhydrous ethanol before calibration. Third, vibration enlarging stitching error across positions, so calibrate under vibration-isolated conditions. On motorized zoom systems every magnification step needs its own calibration; interpolation across steps is not acceptable.