Поверка оптических призм-полигонов: JJG 283 и метод перестановок

Полигоны — эталоны угла; поверка двумя автоколлиматорами методом перестановок даёт погрешности делений.

华旋光电技术部
Поверка оптических призм-полигонов: JJG 283 и метод перестановок

Role and grades of polygons

An optical polygon materializes the full 360° circle as a physical angle standard. Common versions have 8, 12, 24 or 36 faces and are used to calibrate autocollimators, goniometers, dividing heads and rotary tables. Under JJG 283, the verification regulation for regular polygons, the calibration uncertainty of the working angles is classified by grade; a grade-0 polygon typically requires indexing errors within ±1 arc-second and serves to disseminate the angle unit to 0.1 to 0.2 arc-second class instruments. Face flatness and surface roughness (Ra of 0.025 μm or finer) also affect the reflected image and are mandatory items.

Permutation intercomparison

The regulation recommends a full-combination intercomparison with two photoelectric autocollimators: the two instruments face adjacent polygon faces to form a fixed included angle, and readings are taken as the polygon is indexed face by face; the relative instrument positions are then changed and the process repeated. Because the circle closes — the indexing errors must sum to zero by geometry — the n(n-1)/2 independent observations suffice to solve for every face error without any higher-level angle standard. This closure property is the unique advantage of polygon calibration.

Operating essentials

Mount the polygon on an adjustable table and align the rotation axis parallel to the polygon axis to the arc-second level, otherwise tilt errors enter. The autocollimator axis should pass through the center of each working face, coincident with its normal, with eccentricity within 0.1 mm. Control ambient temperature at 20±1 °C with gradients below 0.5 °C/h, shield against air turbulence, and average face readings taken in both approach directions to suppress aiming hysteresis.

Data processing and criteria

Solve the face indexing errors from the intercomparison observations per the regulation formulas, and report the mean, the maximum spread and the expanded uncertainty at k=2. Grade-0 polygons routinely reach 0.1 to 0.2 arc-second expanded uncertainty. A two-year calibration interval is common; a scratched face or damaged coating takes the polygon out of service immediately. For routine checks, a single high-stability autocollimator can be compared against one face; drift beyond 0.3 arc-second justifies recalibration.