XSD-100 Biological Microscope

Biological Microscope

型番XSD-100

    By submitting you agree to our privacy policy.

    01

    製品概要

    XSD-100三目倒置生物显微镜配置大视野目镜,长工作距离平场消色差物镜、长工作距离聚光镜,同时配有相衬装置,可以观察在培养皿和培养瓶中未经染色的透明活体。该仪器特别适用于对活体细胞和组织、流质、沉淀物等进行显微研究,可供科研、高校、医疗、防疫和农牧等部门使用。
    二、 XSD-100三目倒置生物显微镜性能特点:
    1、 配置大视野目镜与长距平场消色差物镜,视场大而且清晰
    2、配置相衬装置,包括长距相衬聚光镜,相衬物镜,对中望远镜
    3、粗微动同轴调焦机构,粗动松紧可调,带锁紧和限位装置,微动格值: 2μm
    4、 6V 20W卤素灯照明,亮度可调
    三、 XSD-100三目倒置生物显微镜性能特点:
    技术参数
    目镜
    大视野 WF10X(直径20mm)
    选购:WF16X(直径11mm)、分划10X(直径20mm),格值0.1mm/格
    物镜
    长距平场消色差物镜 PL L 10X/0.25
    长距平场消色差物镜 PL L 25X/0.40
    长距平场消色差物镜 PL L 40X/0.60
    长距平场消色差相衬物镜 PL L 10X/0.25 PHP
    选配:插板式相衬装置
    物镜:长距平场消色差相衬物镜 PL L 25X/0.40 PHP,PL L 40X/0.60 PHP
    插板式相衬环板:25X、40X
    选配:拉板式相衬装置
    物镜: 长距平场消色差相衬物镜 PL L 10X/0.25 PHP2、PL L 25X/0.40 PHP2、
    PL L 40X/0.60 PHP2
    拉板式相衬环板: 10X、25X、40X
    聚光镜升降可调, NA.=0.4,工作距离:55mm
    目镜筒
    三目镜(倾斜30度,可100%通光摄影)
    调焦机构
    粗微动同轴调焦,粗动松紧可调,带锁紧和限位装置板,微动格值:2μm.
    转换器
    四孔(滚珠内定位)
    选配:五孔(滚珠内定位)
    载物台
    双层机械移动式(尺寸200mmX152mm,移动范围:75mmX30mm)
    标本架
    适合放置直径72mm或77mmX33mm培养器皿
    适合放置直径68mm培养器皿、磨砂玻璃
    滤色片
    蓝滤色片、绿滤色片
    选配:黄滤色片
    插板式相衬装置
    对中望远镜
    10X 插板式相衬环板
    聚光镜升降可调, NA.=0.4,工作距离:30mm
    光源
    6V 20W卤素灯,亮度可调

    XSD-100 Biological Microscope supplied by CHWES. 0 key specifications available on request.

    02

    仕様

    * Generated by rule engine over the product database; for pre-selection only.

    03

    測定原理

    The metallurgical microscope uses an infinity-corrected optical path with reflected Köhler illumination. Light passes the collector, field stop and aperture stop, is deflected by a semi-transparent mirror through the objective onto the specimen, and returns through the same objective to the eyepiece or camera.

    • Bright field for routine microstructure, inclusions and decarburised layers.
    • Dark field collects scattered light to reveal grain boundaries, cracks and non-metallic inclusions.
    • Polarised light identifies anisotropic phases, spheroidised structures and coating stress.
    • DIC converts tiny height differences into contrast, ideal for polished, unetched samples.

    Objectives of 5× to 100× are combined with 10× eyepieces. Higher numerical aperture gives better resolution but shallower depth of field.

    04

    Process animation

    1. Sampling and preparation: sectioning, mounting, grinding, polishing, etching.
    2. Illumination and focus: select BF/DF/POL/DIC and focus.
    3. Image capture: trinocular port, 1× camera adapter, white balance.
    4. Scale calibration: calibrate µm per pixel with a stage micrometer.
    5. Measurement: grain size, case depth, coating thickness, inclusion rating.
    0102030405
    05

    Calculation & formulas

    1. System magnification: M = objective × eyepiece × adapter, e.g. 40× / 10× / 1× = 400×.

    2. Resolution: d = 0.61λ/NA (λ≈550 nm). A 40× objective with NA 0.65 resolves 0.52 µm; an NA 0.95 dry objective reaches 0.35 µm.

    3. Depth of field scales with λ/NA² — higher NA, shallower depth; use low NA for fracture surfaces.

    4. Grain size (GB/T 6394): at 100×, grains N per 645 mm² field follow N = 2G−1; counting 128 grains gives G = log2128 + 1 = 8. Intercept method: mean intercept L̄ = line length ÷ intercepts, G = −6.6437 lg L̄ (mm, 100×).

    5. Effective case depth: distance to 550 HV1, fitted from a micro-hardness traverse and averaged over three positions.

    06

    規格

    StandardTitleScope
    GB/T 13298-2015Metallic materials — examination of microstructureMicrostructure rating
    GB/T 6394-2017Determination of average grain sizeGrain size
    ASTM E3-11Guide for metallographic specimen preparationPreparation
    ASTM E112Determining average grain sizeGrain size
    ISO 643Steels — micrographic determination of grain sizeGrain size

    Export projects often require dual GB and ASTM reports; the software ships both rating charts.

    07

    選定ガイド

    1. Surface or cross-section: stereo microscopes (HX-ZG7, XSZ) for fracture surfaces, solder joints and coatings; upright or inverted metallurgical microscopes (HX-4, HX-G series) for microstructure, case depth and inclusions.

    2. Upright or inverted: upright for small specimens that are flipped often; inverted (HX-G3A/G4C/G5A) for large, heavy parts such as gears, shafts and tube sections that go straight onto the stage without re-mounting.

    3. Optics set the ceiling: plan achromats cover routine rating; semi-apochromats or apochromats with a research stand (HX-G5A) are needed for 1000× precipitates and brightfield/darkfield/polarised switching.

    4. Three levels of digitalisation: a 5 MP camera for images; camera plus analysis software (grain size, case depth, inclusions, porosity) for automatic rating and reports; motorised stage with scanning and stitching for unattended batches.

    5. Do not forget accessories: polariser, darkfield, DIC, and a microhardness interface.

    6. Room: an anti-vibration bench at 20±3 °C for research stands; keep preparation and microscopy rooms separate to avoid abrasive dust.

    SeriesTypeTypical taskRecommended setup
    HX-4A / HX-4BUpright metallurgicalRoutine structure, case depth5×/10×/40× + 5 MP camera
    HX-G2A / G2BInvertedBatch samples, no re-mountingBrightfield + simple polariser
    HX-G3ATrinocular invertedCase depth, rating, reportsCamera + analysis software
    HX-G4C / G5AResearch invertedDarkfield/polarised/DICSemi-apochromat + motorised stage
    HX-ZG7 / XSZ-146StereoFracture, solder, low-mag locatingZoom + ring light

    Common misconceptions: higher magnification is not better (numerical aperture limits resolution, 100× dry is the practical ceiling); ignoring stage travel so large samples do not fit; buying the stand without software and still rating by eye, which kills repeatability.

    08

    試験方法

    Preparation. Section perpendicular to the feature of interest. Keep hot-mounting below 150 °C for aluminium and quenched parts, or use cold mounting. Grind 320→600→800→1200, removing previous scratches at each step; polish with 3 µm then 1 µm diamond. Etchants: 4% nital for carbon steel (3–8 s), aqua regia or electrolytic for stainless, Keller for aluminium.

    Examination. Locate a representative field at 5×/10×, then switch to higher power. Rate grain size to GB/T 6394 by comparison or intercept. Measure case depth at 500× with an ocular scale or calibrated software, averaging five readings.

    Calibration. Verify µm/pixel daily with a stage micrometer, check against a certified specimen quarterly, record three fields per sample and archive images with scale bars.

    Pitfalls. Over-etching widens boundaries and fakes decarburisation; residual scratches can be mistaken for cracks; rounded mounting edges inflate case depth. Keep an unetched specimen for coating and crack review.

    09

    Industry applications

    • Steel and heat treatment: decarburised depth, carburised/nitrided case depth, grain size (GB/T 6394), non-metallic inclusions (GB/T 10561), banding and Widmanstätten rating.
    • Non-ferrous metals: aluminium grain and recrystallisation, alpha/beta ratio in brass, twinning and corrosion morphology in magnesium.
    • Welding and brazing: penetration, HAZ width, porosity and cracks, intermetallic thickness at dissimilar joints.
    • Additive manufacturing: porosity, lack-of-fusion defects, melt-pool morphology in SLM parts, porosity repeatability within 0.2%.
    • Bearings and gears: martensite rating, retained austenite, carbide network and banding, grinding burn and white layer.
    • Electronics: IMC thickness, plating cross-sections, PCB micro-sections, gold wire ball bonds (stereo plus metallurgical).
    • Powders and ceramics: particle morphology and size, sintered porosity, coating cross-section thickness.
    • Failure analysis: fracture surface macro and micro morphology, corrosion products, crack path and origin, with stereo microscopy for low-magnification locating.