W-BB75B Hardness Tester (紫铜)

Hardness Tester · 紫铜

型番W-BB75B
  • 量程0~20HW(相当于20~110HRE)
  • 示值误差0.5HW(5~17HW)
  • 厚 度0.4mm-6mm
  • 管材内径φ10mm以上

By submitting you agree to our privacy policy.

01

製品概要

W-B75b、W-BB75b型仪器分别是W-B75和W-BB75型仪器的细管专用型,仪器砧座减小到Ф5.8mm,可测试内径6mm以上的细管材。
W-B75/BB75系列铜合金韦氏硬度计是轻便的,可以现场快速测试软金属硬度的仪器。测试迅速,简便,一卡即可,硬度值直接读出,符合中国有色标准YS/T471。W-BB75/BB75系列铜合金韦氏硬度计可以测试型材、管材和板材,测试过程对工件无损伤,并且不必取样。特别适于在生产现场、销售现场或施工现场对产品进行快速无损的硬度检查。
W-B75系列韦氏硬度计适于测试黄铜和超硬铝合金材料。
W-BB75系列韦氏硬度计适于测试退火黄铜和紫铜上材料。
标准硬度块经过标准硬度机的检测,附有检测报告。
韦氏硬度计主要用途:
确定工件有无热处理,检查热处理效果,判定工件力学性能是否合格。
确定工件是否为不适当的合金加工而成,判定材料合金成份是否合格。
测试不便送到实验室的过长、过重工件或装配件。
区分材料的硬度级别。判断材料的退火、1/8硬、1/4硬、1/2硬、全硬等各种热处理状态。
技术参数:
量程:0~20HW(相当于20~110HRE);
示值误差:0.5HW(5~17HW);
厚 度:0.4mm-6mm;
管材内径:φ10mm以上;
重量:0.5kg;
标准配置:
主机;
韦氏硬度块;
备用压针;
校正扳手;
小螺丝刀;
仪器箱;
可选附件:
压针;
韦氏硬度块;
表窗玻璃;
应用说明:
W-20型:通用型,用于普通铝型材;
W-20a型:用于厚度小于13mm的铝型材;
W-20b型:用于内径大于6mm的铝管材;
W-B75型:用于黄铜管、黄铜带;
W-BB75型:用于紫铜管、紫铜带;
W-B92型:用于软态不锈钢、冷轧钢带;
仪器选型指南:
铝合金:
硬度范围:25~110HRE;58~131HV;
W-20工件尺寸mm:厚度0.6-6,内径>10;
W-20A工件尺寸mm:厚度0.6-13,内径>10;
W-20B工件尺寸mm:厚度0.6-8,内径>6;
黄铜:
硬度范围:63~105HRF;
W-B75工件尺寸mm:厚度0.6-6,内径>10;
W-B75B工件尺寸mm:厚度0.6-8,内径>6;
紫铜:
硬度范围:18~100HRE;
W-BB75工件尺寸mm:厚度0.6-6,内径>10;
W-BB75B工件尺寸mm:厚度0.6-8,内径>6;
不锈钢带冷轧薄钢带:
硬度范围:50~92HRB;
W-B92工件尺寸mm:厚度0.4-6,内径>10;

W-BB75B Hardness Tester (紫铜) supplied by CHWES. 5 key specifications available on request.

02

仕様

量程0~20HW(相当于20~110HRE)
示值误差0.5HW(5~17HW)
厚 度0.4mm-6mm
管材内径φ10mm以上
重量0.5kg

* Generated by rule engine over the product database; for pre-selection only.

03

測定原理

Hardness testing presses an indenter into the surface under a constant load and converts indentation size or depth into a hardness value. Vickers uses a 136° diamond pyramid: HV = 0.1891 × F(N)/d²(mm), where d is the mean diagonal.

Hot hardness is performed in vacuum or high-purity inert gas. Specimen and indenter are heated independently and must be at the same temperature or corrected per the standard; loading, dwell and unloading all happen inside the hot zone, and the indentation is measured through a high-temperature optic or after cooling. The HMAS series uses a uniform hot zone with ±5 °C uniformity from 200 to 1600 °C.

  • Vickers 0.01–100 kgf: full range, thin layers and case-depth gradients.
  • Rockwell C 150 kgf: fast batch checking of hardened parts.
  • Brinell: castings, annealed and coarse-grain materials.
  • Leeb: portable rebound testing on large parts.
04

Process animation

  1. Preparation: polish to Ra ≤ 0.4 µm, remove work-hardened layers.
  2. Mount and align: place the specimen, position the indenter over the test point.
  3. Heat up (hot hardness): evacuate to ≤10⁻² Pa or backfill argon, ramp and soak.
  4. Load: apply force at the set rate, dwell 10–15 s.
  5. Measure: read diagonals or depth, compute HV/HRC.
  6. Gradient: step along depth and plot hardness versus distance.
0102030405
05

Calculation & formulas

1. Rockwell: HR = K − h/0.002 (K = 100 for HRC, 130 for HRB). At h = 0.078 mm, HRC = 61.

2. Brinell: HBW = 2F/(πD(D−√(D²−d²))); with F=1838.6 N, D=2.5 mm, d=1.1 mm, HBW ≈ 192.

3. Vickers: HV = 0.1891·F/d² (F in N); F=9.807 N with d=0.05 mm gives HV ≈ 742.

4. Test forces: choose Brinell loads by 0.045F/D²; Rockwell 10 kgf preliminary with 60/100/150 kgf total; surface scales 3 / 15/30/45 kgf.

5. Conversions: HRC ≈ HBW/10 only as a rough mid-range guide — official reports must follow the GB/T 1172 tables.

06

規格

StandardTitleScope
GB/T 4340.1-2009Vickers hardness test — test methodVickers
GB/T 230.1-2018Rockwell hardness testRockwell
GB/T 231.1-2018Brinell hardness testBrinell
ISO 6507-1Vickers hardness testExport
ASTM E92Vickers hardness of metallic materialsAmericas
GB/T 17394.1-2014Leeb hardness testPortable

Hot hardness has no unified national standard; ASTM E92 practice is normally cited with temperature, atmosphere, dwell and indenter material stated in the report.

07

選定ガイド

1. Principle follows the target: Rockwell for macro hardness and heat-treatment acceptance; Brinell for cast iron, annealed and forged parts where a large indent is representative; Vickers or micro-Vickers for thin layers, coatings and small areas.

2. Bench or portable: bench units (HRC-150SMJ, HBRVS-187.5) where parts can be brought in; portable Leeb for large moulds, in-service piping and non-removable parts, corrected against a bench reference, typically within ±2 HRC.

3. One for many: mixed, low-volume shops do well with a Brinell-Vickers-Rockwell combination unit covering HRA/HRB/HRC/HV/HB.

4. Automation: manual below 100 parts per operator per day; motorised elevation, automatic loading and data upload (HRC-150SMJ) above 200 or for unattended runs.

5. Reference blocks: two to three blocks covering the working range, checked every shift and sent for annual calibration.

6. Specimen requirements: Rockwell needs thickness at least ten times the indent depth and Ra ≤1.6 µm; otherwise switch to Vickers or superficial Rockwell.

ModelPrincipleRangeTypical use
HRC-150SMJRockwell, motorised digitalHRA/HRB/HRCBatch sampling in heat treatment
HRC-150SM / HRC-150MRockwellHRA/HRB/HRCIncoming and in-process checks
HRC-45MSuperficial Rockwell15N/30N/45NSheet, plating, case layers
HRS-150M / XHRS-150MDigital RockwellHRA/HRB/HRCLines needing data upload
HBRVS-187.5Brinell-Vickers-RockwellHBW/HRA/HRB/HRC/HVMixed low-volume production

Three rules for trustworthy data: check the reference block every shift and record it; keep the support face flat and clean; keep indent spacing at least three indent diameters.

08

試験方法

Specimen. Test and support faces parallel; thickness at least 1.5 × diagonal (Vickers) or 10 × depth (Rockwell); Ra ≤ 0.4 µm, free of scale and oil.

Room temperature. Select load, apply within 10–15 s, dwell 10–15 s, unload, measure diagonals. Spacing between indentations ≥3d, from the edge ≥2.5d.

Hot hardness. Load the specimen, evacuate or backfill argon, ramp and soak 20–30 min, lower the indenter, load and dwell, unload, cool or measure hot, then apply the temperature correction.

Gradient. Step every 0.05–0.1 mm to the hardness plateau; determine effective case depth at 550 HV1 per GB/T 9450.

Quality control. Check a certified block (e.g. 700 HV1) each shift, recalibrate beyond ±3%, and log furnace curves and atmosphere for hot tests.

09

Industry applications

  • Heat-treatment shops: Rockwell HRC sampling of quenched and tempered parts, carburised surface hardness, Brinell HBW on annealed stock.
  • Castings and forgings: body hardness and HAZ values, section hardness gradients indicating cooling rate and structural uniformity.
  • Moulds and tools: HRC versus wear resistance after hardening and tempering, edge-to-core hardness difference on cutting tools.
  • Thin and small parts: case-hardened layers, coatings and sheet need Vickers or superficial Rockwell (HBRVS-187.5 covers HRA/HRB/HRC/HV/HB).
  • Bearings and gears: ring and raceway hardness, case-to-core difference against carburising requirements.
  • Pressure vessels and piping: weld and HAZ hardness to detect brittle microstructures, typically ≤350 HV10.
  • Incoming verification: hardness as the fastest screening method when material grades are mixed, confirmed by OES.
  • Field and in-service: portable Leeb or ultrasonic hardness testers on large immovable parts, corrected against a bench reference.
10

Price and lead time for this model

By submitting you agree to our privacy policy.

⇄0比較仕様比較