HRS-150 Rockwell Hardness Tester

Rockwell Hardness Tester

型番HRS-150
  • 洛氏标尺HRA、HRB、HRC、HRD、HRE、HRF、HRG、HRH、HRK
  • 初试验力98.07N (10kgf)
  • 总试验力588N(60kgf),980N(100kgf),1471N(150kgf)
  • 读数方式大屏幕数字显示,尺寸:100X77mm

By submitting you agree to our privacy policy.

01

製品概要

HRS-150数显洛氏硬度计:属于自动型洛氏硬度计,外观新颖大屏幕显示,菜单式结构,采用微机控制,硬度值以数字直接显示.硬度测试结果可以直接打印输出,实现了自动化,消除了操作和读数误差。还具有硬度之间的硬度换算等功能。适用于金属材料的洛氏硬度试验。本系列洛氏硬度计操作简便迅速,性能稳定,使用广泛。可进行黑色金属和有色金属的洛氏硬度测定,广泛适用于工厂车间和计量部门.
与其它家区别:
1、90%以上工件自主加工,而非外协;
2、机壳铸铁铸造而非焊接,壳体重量达70公斤;
3、整机重量达110公斤;
4、整机基本上为金属零件;
5、壳体壁厚为14-16mm;
6、自主油漆喷涂技术,高档金属漆;
产品特点:
1、液晶屏幕显示、菜单式结构,采用微机控制,硬度值以数字直接显示;
2、铸铁机壳,稳固、耐用、无变形、承受力大;
3、壳体壁厚达14-16mm,稳固、无变形、抗震性好;
4、自主加工零件,可替换性高,后续零件供应稳定;
5、自主油漆喷涂技术,任意颜色可选择,能够美化和统一实验室色彩;
技术参数:
操作方式
数显
洛氏标尺
HRA、HRB、HRC、HRD、HRE、HRF、HRG、HRH、HRK
初试验力
98.07N (10kgf)
总试验力
588N(60kgf),980N(100kgf),1471N(150kgf)
读数方式
大屏幕数字显示,尺寸:100X77mm
硬度分辨率
0.1HR
试件最大高度
180mm
压头中心距离
160mm
数据输出
内置微型打印机
试验力保持时间(S):
机器设定0~99秒,最小变动量1秒
转换标尺
表面洛氏、布氏、维氏
硬度测量范围
HRA:20~88 ,HRB:20~100 ,HRC:20~70
执行国家标准
GB/T230.2国家标准,JJG112检定规程
丝杆直径
外径Ф42mm
大平台直径
Ф150mm
中平台直径
Ф60mm,厚21.5mm
V型台尺寸
Ф40mm,开口31.5mm
壳体工艺及壁厚
铸造壳体,壁厚14-16mm
外形尺寸
535×150×700mm
电源
220V
重量
80Kg

HRS-150 Rockwell Hardness Tester supplied by CHWES. 20 key specifications available on request.

02

仕様

洛氏标尺HRA、HRB、HRC、HRD、HRE、HRF、HRG、HRH、HRK
初试验力98.07N (10kgf)
总试验力588N(60kgf),980N(100kgf),1471N(150kgf)
读数方式大屏幕数字显示,尺寸:100X77mm
硬度分辨率0.1HR
试件最大高度180mm
压头中心距离160mm
数据输出内置微型打印机
试验力保持时间(S):机器设定0~99秒,最小变动量1秒
转换标尺表面洛氏、布氏、维氏
硬度测量范围HRA:20~88 ,HRB:20~100 ,HRC:20~70
执行国家标准GB/T230.2国家标准,JJG112检定规程
丝杆直径外径Ф42mm
大平台直径Ф150mm
中平台直径Ф60mm,厚21.5mm
V型台尺寸Ф40mm,开口31.5mm
壳体工艺及壁厚铸造壳体,壁厚14-16mm
外形尺寸535×150×700mm
电源220V
重量80Kg

* Generated by rule engine over the product database; for pre-selection only.

03

測定原理

Hardness testing presses an indenter into the surface under a constant load and converts indentation size or depth into a hardness value. Vickers uses a 136° diamond pyramid: HV = 0.1891 × F(N)/d²(mm), where d is the mean diagonal.

Hot hardness is performed in vacuum or high-purity inert gas. Specimen and indenter are heated independently and must be at the same temperature or corrected per the standard; loading, dwell and unloading all happen inside the hot zone, and the indentation is measured through a high-temperature optic or after cooling. The HMAS series uses a uniform hot zone with ±5 °C uniformity from 200 to 1600 °C.

  • Vickers 0.01–100 kgf: full range, thin layers and case-depth gradients.
  • Rockwell C 150 kgf: fast batch checking of hardened parts.
  • Brinell: castings, annealed and coarse-grain materials.
  • Leeb: portable rebound testing on large parts.
04

Process animation

  1. Preparation: polish to Ra ≤ 0.4 µm, remove work-hardened layers.
  2. Mount and align: place the specimen, position the indenter over the test point.
  3. Heat up (hot hardness): evacuate to ≤10⁻² Pa or backfill argon, ramp and soak.
  4. Load: apply force at the set rate, dwell 10–15 s.
  5. Measure: read diagonals or depth, compute HV/HRC.
  6. Gradient: step along depth and plot hardness versus distance.
0102030405
05

Calculation & formulas

1. Rockwell: HR = K − h/0.002 (K = 100 for HRC, 130 for HRB). At h = 0.078 mm, HRC = 61.

2. Brinell: HBW = 2F/(πD(D−√(D²−d²))); with F=1838.6 N, D=2.5 mm, d=1.1 mm, HBW ≈ 192.

3. Vickers: HV = 0.1891·F/d² (F in N); F=9.807 N with d=0.05 mm gives HV ≈ 742.

4. Test forces: choose Brinell loads by 0.045F/D²; Rockwell 10 kgf preliminary with 60/100/150 kgf total; surface scales 3 / 15/30/45 kgf.

5. Conversions: HRC ≈ HBW/10 only as a rough mid-range guide — official reports must follow the GB/T 1172 tables.

06

規格

StandardTitleScope
GB/T 4340.1-2009Vickers hardness test — test methodVickers
GB/T 230.1-2018Rockwell hardness testRockwell
GB/T 231.1-2018Brinell hardness testBrinell
ISO 6507-1Vickers hardness testExport
ASTM E92Vickers hardness of metallic materialsAmericas
GB/T 17394.1-2014Leeb hardness testPortable

Hot hardness has no unified national standard; ASTM E92 practice is normally cited with temperature, atmosphere, dwell and indenter material stated in the report.

07

選定ガイド

1. Principle follows the target: Rockwell for macro hardness and heat-treatment acceptance; Brinell for cast iron, annealed and forged parts where a large indent is representative; Vickers or micro-Vickers for thin layers, coatings and small areas.

2. Bench or portable: bench units (HRC-150SMJ, HBRVS-187.5) where parts can be brought in; portable Leeb for large moulds, in-service piping and non-removable parts, corrected against a bench reference, typically within ±2 HRC.

3. One for many: mixed, low-volume shops do well with a Brinell-Vickers-Rockwell combination unit covering HRA/HRB/HRC/HV/HB.

4. Automation: manual below 100 parts per operator per day; motorised elevation, automatic loading and data upload (HRC-150SMJ) above 200 or for unattended runs.

5. Reference blocks: two to three blocks covering the working range, checked every shift and sent for annual calibration.

6. Specimen requirements: Rockwell needs thickness at least ten times the indent depth and Ra ≤1.6 µm; otherwise switch to Vickers or superficial Rockwell.

ModelPrincipleRangeTypical use
HRC-150SMJRockwell, motorised digitalHRA/HRB/HRCBatch sampling in heat treatment
HRC-150SM / HRC-150MRockwellHRA/HRB/HRCIncoming and in-process checks
HRC-45MSuperficial Rockwell15N/30N/45NSheet, plating, case layers
HRS-150M / XHRS-150MDigital RockwellHRA/HRB/HRCLines needing data upload
HBRVS-187.5Brinell-Vickers-RockwellHBW/HRA/HRB/HRC/HVMixed low-volume production

Three rules for trustworthy data: check the reference block every shift and record it; keep the support face flat and clean; keep indent spacing at least three indent diameters.

08

試験方法

Specimen. Test and support faces parallel; thickness at least 1.5 × diagonal (Vickers) or 10 × depth (Rockwell); Ra ≤ 0.4 µm, free of scale and oil.

Room temperature. Select load, apply within 10–15 s, dwell 10–15 s, unload, measure diagonals. Spacing between indentations ≥3d, from the edge ≥2.5d.

Hot hardness. Load the specimen, evacuate or backfill argon, ramp and soak 20–30 min, lower the indenter, load and dwell, unload, cool or measure hot, then apply the temperature correction.

Gradient. Step every 0.05–0.1 mm to the hardness plateau; determine effective case depth at 550 HV1 per GB/T 9450.

Quality control. Check a certified block (e.g. 700 HV1) each shift, recalibrate beyond ±3%, and log furnace curves and atmosphere for hot tests.

09

Industry applications

  • Heat-treatment shops: Rockwell HRC sampling of quenched and tempered parts, carburised surface hardness, Brinell HBW on annealed stock.
  • Castings and forgings: body hardness and HAZ values, section hardness gradients indicating cooling rate and structural uniformity.
  • Moulds and tools: HRC versus wear resistance after hardening and tempering, edge-to-core hardness difference on cutting tools.
  • Thin and small parts: case-hardened layers, coatings and sheet need Vickers or superficial Rockwell (HBRVS-187.5 covers HRA/HRB/HRC/HV/HB).
  • Bearings and gears: ring and raceway hardness, case-to-core difference against carburising requirements.
  • Pressure vessels and piping: weld and HAZ hardness to detect brittle microstructures, typically ≤350 HV10.
  • Incoming verification: hardness as the fastest screening method when material grades are mixed, confirmed by OES.
  • Field and in-service: portable Leeb or ultrasonic hardness testers on large immovable parts, corrected against a bench reference.
10

Price and lead time for this model

By submitting you agree to our privacy policy.

⇄0比較仕様比較